F&J Model D-814 Air Flow Calibrator |
|
|
 |
 |
MEASUREMENT PRINCIPLE Flow Sensor: Venturi Tube Standardization: Correction for standard temperature and barometric pressure Curve linearization: Individually calibrated and software corrected
MEASUREMENT RANGES Range of 2 - 14 CFM (56 - 400 LPM) Temperature: -4° F - +122°F (-20°C – +50° C) Barometric pressure: 30 – 24 In-Hg (760 – 609 mm-Hg) approx. Sea level to 6000 ft. elevation above sea level
MEASUREMENT ACCURACY Air flow: +/- 2% of full scale (Optional +/- 1% of full scale) Temperature (Typical): 1°C over range -4° F - +122°F (-20°C – +50° C) Barometric pressure: +/- 1% over measurement range
DISPLAY 6 Digit LCD 0.4" high
GENERAL INFORMATION
|
Power requirements: |
100–120 VAC;50/60Hz |
|
|
Operating temperature: |
-4° F - +122°F |
(-20°C – +50° C) |
|
Storage temperature: |
-20° F - +122°F |
-29°C – +70° C) |
|
Dimension (L×W×H) |
9.75"×8.25"×10.75" |
(248×210×273mm) |
|
Weight: |
8 lbs. 6 oz. |
(3.8 Kgs.) |
|
Installation Category: |
Pollution Degree 2 |
|
|
Enclosure Rating: |
IPXO |
|
|
|
 |
Related Items |
|
|
|
|
|
|
|
|
|